Applications of UV–Visible spectroscopy in single and multicomponent analysis MCQs With Answer

Introduction: UV–Visible spectroscopy (UV-Vis) is an essential tool in pharmaceutical analysis for qualitative and quantitative determination of drugs. This technique relies on the Beer-Lambert law, absorbance at characteristic λmax, and parameters like molar absorptivity and path length to perform accurate single-component assays and robust multicomponent analysis. In pharmaceuticals, methods such as the simultaneous equation method, derivative spectrophotometry, ratio spectra and multivariate calibration enable resolution of overlapping spectra in combined formulations. Understanding sample preparation, solvent effects, linearity, sensitivity (LOD/LOQ), and method validation is crucial for reliable UV-Vis assays. Now let’s test your knowledge with 30 MCQs on this topic.

Q1. What fundamental law relates absorbance to concentration in UV–Visible spectroscopy?

  • Beer-Lambert law
  • Fick’s law
  • Henry’s law
  • Hooke’s law

Correct Answer: Beer-Lambert law

Q2. In single-component analysis by UV-Vis, the wavelength of maximum absorbance is referred to as:

  • Isoelectric point
  • λmax
  • Chromatographic peak
  • Retention time

Correct Answer: λmax

Q3. Which parameter expresses how strongly a compound absorbs light at a given wavelength?

  • pKa
  • Molar absorptivity (ε)
  • Retention factor (Rf)
  • Viscosity

Correct Answer: Molar absorptivity (ε)

Q4. When two components in a mixture absorb at overlapping wavelengths, which UV-Vis method can resolve their concentrations?

  • Simultaneous equation method
  • Distillation
  • Gravimetric analysis
  • pH titration

Correct Answer: Simultaneous equation method

Q5. In the simultaneous equation method for two components A and B, what two measurements are required?

  • Absorbances at two selected wavelengths
  • Masses of A and B
  • Retention times in HPLC
  • pH and temperature

Correct Answer: Absorbances at two selected wavelengths

Q6. Which technique enhances resolution of overlapping spectra by computing first or higher derivatives of the absorption spectrum?

  • Derivative spectrophotometry
  • Fluorimetry
  • Mass spectrometry
  • NMR spectroscopy

Correct Answer: Derivative spectrophotometry

Q7. What is the significance of an isosbestic point in a binary mixture?

  • Absorbance is independent of the relative concentrations of the two species
  • It indicates pH instability
  • It marks the chromatographic retention time
  • It is used to measure viscosity

Correct Answer: Absorbance is independent of the relative concentrations of the two species

Q8. Which factor does NOT directly affect absorbance in UV-Vis measurements?

  • Path length of the cuvette
  • Concentration of analyte
  • Molar absorptivity
  • Magnetic field strength

Correct Answer: Magnetic field strength

Q9. The absorbance additivity principle states that:

  • Total absorbance equals the sum of individual component absorbances at a given wavelength
  • Absorbance multiplies for mixtures
  • Only the most absorbing species contributes
  • Absorbance is independent of concentration

Correct Answer: Total absorbance equals the sum of individual component absorbances at a given wavelength

Q10. Which solvent parameter can significantly shift λmax of a drug during UV-Vis analysis?

  • Dielectric constant (solvent polarity)
  • Magnetic susceptibility
  • Boiling point of solvent
  • Density of container glass

Correct Answer: Dielectric constant (solvent polarity)

Q11. For method validation in UV-Vis assays, which parameter assesses the smallest reliably detectable concentration?

  • Limit of detection (LOD)
  • Linearity range
  • Precision
  • Specificity

Correct Answer: Limit of detection (LOD)

Q12. Which approach uses a ratio of absorbances at two wavelengths to quantify a component in a mixture?

  • Q-analysis (absorbance ratio method)
  • Gravimetric titration
  • Thermogravimetry
  • Conductometry

Correct Answer: Q-analysis (absorbance ratio method)

Q13. In multicomponent UV-Vis analysis, multivariate calibration (e.g., PLS) is advantageous because it:

  • Handles highly overlapping spectra and correlates many wavelengths simultaneously
  • Requires no instrument calibration
  • Eliminates the need for standards
  • Operates only at single wavelength

Correct Answer: Handles highly overlapping spectra and correlates many wavelengths simultaneously

Q14. Which instrumental parameter determines spectral resolution and influences peak shape?

  • Slit width
  • Sample viscosity
  • Laboratory humidity
  • Manufacturer logo

Correct Answer: Slit width

Q15. When constructing a calibration curve for a single component, which plot is most common?

  • Absorbance versus concentration
  • Time versus temperature
  • Mass versus volume
  • pH versus conductivity

Correct Answer: Absorbance versus concentration

Q16. Which of the following is a common interference in UV-Vis pharmaceutical assays?

  • Excipients that absorb at similar wavelengths
  • Ambient sound
  • Colour of the laboratory bench
  • Time of day

Correct Answer: Excipients that absorb at similar wavelengths

Q17. Ratio spectra derivative spectrophotometry is used primarily to:

  • Resolve overlapping signals in binary and ternary mixtures
  • Measure boiling points
  • Separate enantiomers
  • Sterilize samples

Correct Answer: Resolve overlapping signals in binary and ternary mixtures

Q18. For a fixed path length, doubling the concentration of an absorbing species will:

  • Double the absorbance (within linear range)
  • Halve the absorbance
  • Increase absorbance by four times
  • Have no effect on absorbance

Correct Answer: Double the absorbance (within linear range)

Q19. What is the main purpose of using a blank in UV-Vis measurements?

  • To correct for solvent and cuvette absorbance
  • To increase instrument throughput
  • To sterilize the spectrophotometer
  • To calibrate temperature

Correct Answer: To correct for solvent and cuvette absorbance

Q20. Which quality parameter evaluates closeness of measured values to the true value in UV-Vis assays?

  • Accuracy
  • Repeatability
  • Robustness
  • Sensitivity

Correct Answer: Accuracy

Q21. When selecting wavelengths for simultaneous equation method, you should choose wavelengths where:

  • Each analyte has significant and differing absorbance contributions
  • Both analytes have zero absorbance
  • Only one analyte absorbs at both wavelengths
  • The solvent absorbs maximally

Correct Answer: Each analyte has significant and differing absorbance contributions

Q22. What is the effect of stray light on UV-Vis measurements?

  • It causes deviation from linearity at high absorbance values
  • It increases sample concentration
  • It reduces solvent polarity
  • It stabilizes λmax

Correct Answer: It causes deviation from linearity at high absorbance values

Q23. Which sample handling practice improves accuracy of UV-Vis assays?

  • Using matched cuvettes and consistent path length
  • Changing cuvettes between each measurement without rinsing
  • Using colored cuvettes to mask impurities
  • Measuring without blanks

Correct Answer: Using matched cuvettes and consistent path length

Q24. In a ternary mixture, which UV-Vis approach can help quantify three overlapping components?

  • Multivariate calibration or multi-wavelength methods
  • Single-point titration
  • Paper chromatography
  • Simple dilution without measurement

Correct Answer: Multivariate calibration or multi-wavelength methods

Q25. Which pharmacopoeial requirement is commonly assessed for UV-Vis analytical methods?

  • System suitability, linearity, accuracy, precision, specificity
  • Boiling point range only
  • Colour of reagent bottles
  • Manufacturer’s logo compliance

Correct Answer: System suitability, linearity, accuracy, precision, specificity

Q26. The term “specificity” in UV-Vis method validation means:

  • The method can measure the analyte in presence of other components
  • The method is cheap
  • The method uses the latest instrument
  • The method is complicated

Correct Answer: The method can measure the analyte in presence of other components

Q27. When would you apply baseline correction during spectral acquisition?

  • To remove background drift and instrument noise
  • To increase sample concentration
  • To change the chemical structure
  • To adjust sample pH

Correct Answer: To remove background drift and instrument noise

Q28. What does a high molar absorptivity (ε) indicate about an analyte?

  • It has strong absorbance per molar concentration and is highly sensitive in UV-Vis
  • It has low solubility
  • It is nonpolar
  • It is volatile

Correct Answer: It has strong absorbance per molar concentration and is highly sensitive in UV-Vis

Q29. Which practice helps minimize matrix effects in multicomponent UV-Vis assays of formulations?

  • Appropriate sample dilution, extraction, and use of matched placebos
  • Measuring undiluted suspensions without filtration
  • Using tap water as solvent
  • Avoiding use of standards

Correct Answer: Appropriate sample dilution, extraction, and use of matched placebos

Q30. Which statement best describes the use of UV-Vis spectroscopy in pharmaceutical quality control?

  • It provides rapid, cost-effective assays for drug identity, assay, and dissolution when properly validated
  • It always replaces HPLC in all assays
  • It cannot be used for multicomponent formulations
  • It is unsuitable for routine analysis

Correct Answer: It provides rapid, cost-effective assays for drug identity, assay, and dissolution when properly validated

Leave a Comment