Dilution Using Absorbance
This calculator helps you determine the volumes needed to dilute a stock solution to a desired final absorbance, utilizing the principles of the Beer-Lambert Law.
The Beer-Lambert Law states that absorbance (A) is directly proportional to concentration (c) for a given substance at a specific wavelength and path length (A = εlc). Because absorbance is directly proportional to concentration, we can substitute absorbance values directly into the standard dilution equation M₁V₁ = M₂V₂ (or C₁V₁ = C₂V₂).
A₁V₁ = A₂V₂
Where:
• A₁ = Absorbance of the stock solution.
• V₁ = Volume of the stock solution needed (this is what we solve for).
• A₂ = Desired absorbance of the final diluted solution.
• V₂ = Desired final volume of the diluted solution.
By rearranging this formula to V₁ = (A₂ * V₂) / A₁, we can calculate the required volume of the stock solution. The volume of diluent (e.g., buffer or solvent) needed is then simply the final volume minus the stock volume: Volume of Diluent = V₂ - V₁.
Frequently Asked Questions
What is the ideal absorbance range for accuracy?
Most spectrophotometers provide the most accurate and reliable readings within an absorbance range of approximately 0.1 to 1.0.
• Readings below 0.1 can be significantly affected by instrument noise and stray light, leading to poor precision.
• Readings above 1.0 (especially > 1.5 or 2.0 depending on the instrument) often deviate from the linear relationship predicted by the Beer-Lambert Law due to factors like high concentration effects, stray light, or detector limitations.
Therefore, it's often necessary to dilute samples so their absorbance falls within this optimal range for accurate quantification.
Why is a 'blank' solution necessary?
A "blank" solution contains everything present in your sample solution *except* the substance you are trying to measure (the analyte). This typically includes the solvent or buffer. Measuring the absorbance of the blank first allows you to zero the spectrophotometer. This effectively subtracts any absorbance caused by the cuvette itself or the solvent/buffer, ensuring that the final absorbance reading you obtain for your sample is due only to the analyte of interest.
Does this work at any wavelength?
Yes, the principle A₁V₁ = A₂V₂ holds true as long as both absorbance values (A₁ and A₂) are measured at the same wavelength and under the same conditions (e.g., same cuvette path length, same solvent). The specific wavelength chosen usually corresponds to the maximum absorbance (λmax) of the substance being measured to maximize sensitivity.