Qualification of tunnels and dry heat sterilizers MCQs With Answer

Qualification of tunnels and dry heat sterilizers is a critical topic in pharmaceutical validation, ensuring that thermal sterilization equipment consistently achieves required microbial inactivation without compromising product quality. This blog quiz is tailored for M.Pharm students to deepen understanding of installation, operational and performance qualifications (IQ/OQ/PQ), thermal mapping, instrumentation, biological indicators, load effects, and acceptance criteria. Questions emphasize practical considerations such as sensor placement, worst‑case load scenarios, calibration standards, and interpretation of thermal profiles. Use this set to reinforce theory, prepare for viva/board exams, and apply validation principles to real-world sterilization systems in pharmaceutical manufacturing.

Q1. Which biological indicator organism is most appropriate for validating a dry heat sterilization process?

  • Bacillus stearothermophilus
  • Geobacillus thermophilus
  • Bacillus atrophaeus
  • Escherichia coli

Correct Answer: Bacillus atrophaeus

Q2. During IQ of a tunnel sterilizer, which document is least likely to be verified?

  • Manufacturer’s drawings and serial number
  • Calibration certificates for installed thermocouples
  • Microbial growth curve for production batches
  • Electrical and utility connections

Correct Answer: Microbial growth curve for production batches

Q3. The primary objective of Operational Qualification (OQ) for a dry heat sterilizer is to:

  • Install the equipment according to manufacturer instructions
  • Demonstrate the equipment operates within defined limits under simulated conditions
  • Prove consistent sterilization on full production loads
  • Qualify the production personnel handling the sterilizer

Correct Answer: Demonstrate the equipment operates within defined limits under simulated conditions

Q4. Which sensor placement strategy is most appropriate for thermal mapping of a tunnel sterilizer?

  • Place all sensors only at the air inlet
  • Place sensors at representative product positions, extremes, and along conveyor direction
  • Place sensors only on the chamber walls
  • Use a single sensor in the warmest location

Correct Answer: Place sensors at representative product positions, extremes, and along conveyor direction

Q5. For dry heat sterilization of glassware, a commonly recommended cycle is approximately:

  • 121°C for 15 minutes
  • 160°C for 2 hours
  • 100°C for 30 minutes
  • 75°C for 10 minutes

Correct Answer: 160°C for 2 hours

Q6. Which of the following is a critical operational parameter to monitor in a hot-air tunnel sterilizer?

  • Conveyor speed (dwell time)
  • pH of cleaning solution
  • Volume of packaging material shipped
  • Electrical conductivity of water

Correct Answer: Conveyor speed (dwell time)

Q7. In qualification terminology, PQ (Performance Qualification) primarily involves:

  • Verifying installation drawings and spare parts list
  • Demonstrating consistent performance under normal production conditions using actual product loads
  • Calibrating instruments used for mapping
  • Writing the equipment SOP

Correct Answer: Demonstrating consistent performance under normal production conditions using actual product loads

Q8. Which of the following best describes a worst-case load for dry heat sterilizer PQ?

  • An empty chamber with maximum airflow
  • A heavily loaded chamber with dense, insulating items and worst thermal inertia
  • A chamber loaded with metal trays only
  • A chamber with only biological indicators exposed to air

Correct Answer: A heavily loaded chamber with dense, insulating items and worst thermal inertia

Q9. Acceptance criteria for temperature uniformity during thermal mapping are typically stated as:

  • Any temperature is acceptable as long as the setpoint is reached once
  • Temperatures must remain within a specified range (e.g., ±2°C) across mapped locations
  • Only the average temperature matters, not local variations
  • Temperature variation must be within ±20°C

Correct Answer: Temperatures must remain within a specified range (e.g., ±2°C) across mapped locations

Q10. Which calibration practice is essential before performing OQ/PQ thermal mapping?

  • Calibrating air filters
  • Calibration and accuracy verification of thermocouples/data loggers with traceable standards
  • Calibrating the conveyor motor speed by feel
  • No calibration is needed if manufacturer supplies sensors

Correct Answer: Calibration and accuracy verification of thermocouples/data loggers with traceable standards

Q11. The primary heat transfer mechanism in forced‑air dry heat sterilizers is:

  • Condensation
  • Conduction through water
  • Convection by circulated hot air
  • Ultraviolet radiation

Correct Answer: Convection by circulated hot air

Q12. During qualification of a tunnel sterilizer, product temperature lag compared to air temperature is important because:

  • It indicates the sterilizer has excess power
  • Product core may not reach required sterilizing temperature, creating a cold spot risk
  • It shows the thermocouples are faulty
  • It is irrelevant for surface-sterilized packaging

Correct Answer: Product core may not reach required sterilizing temperature, creating a cold spot risk

Q13. When using biological indicators in PQ for a dry heat sterilizer, the correct recovery step is to:

  • Immediately discard the BI after the cycle
  • Incubate the BI in appropriate growth medium under specified conditions to detect survivors
  • Rinse the BI with sterile water and reuse
  • Expose the BI to UV light and observe

Correct Answer: Incubate the BI in appropriate growth medium under specified conditions to detect survivors

Q14. Which qualification activity verifies that tunnel conveyor speed settings produce the intended dwell time at set temperature zones?

  • IQ documentation review
  • Conveyor speed validation during OQ with timing measurements and thermoprobes
  • Electrical load testing
  • Cleaning validation

Correct Answer: Conveyor speed validation during OQ with timing measurements and thermoprobes

Q15. A common acceptance criterion for lethality demonstration in dry heat sterilization is achieving a specified log reduction. Which term denotes a one‑log (90%) reduction in a microbial population?

  • D-value
  • Z-value
  • F0-value
  • G-value

Correct Answer: D-value

Q16. Which parameter describes the temperature change required to change the D‑value by a factor of ten?

  • D-value
  • Z-value
  • F-value
  • k-value

Correct Answer: Z-value

Q17. During PQ of a tunnel sterilizer handling prefilled syringes, what is a key concern to monitor besides temperature?

  • Sterilizer paint color
  • Product integrity: deformation, seal integrity, and content stability
  • Manufacturer’s country of origin
  • Ambient noise levels

Correct Answer: Product integrity: deformation, seal integrity, and content stability

Q18. If thermal mapping shows a persistent cold spot in a tunnel sterilizer, the appropriate corrective action is to:

  • Ignore it if most locations pass
  • Adjust process parameters (temperature, airflow, dwell time) or modify load/configuration and revalidate
  • Reduce production volume without investigation
  • Replace all thermocouples immediately

Correct Answer: Adjust process parameters (temperature, airflow, dwell time) or modify load/configuration and revalidate

Q19. Which regulatory expectation applies to documentation produced during qualification of sterilization equipment?

  • Documents should be informal and handwritten only
  • Complete, traceable, version‑controlled records including protocols, raw data, deviations, and final reports
  • No documentation is required for legacy equipment
  • Only verbal approvals are sufficient

Correct Answer: Complete, traceable, version‑controlled records including protocols, raw data, deviations, and final reports

Q20. Which statement best describes the role of simulated worst‑case runs in OQ for tunnels and dry heat sterilizers?

  • They are only needed after a failure occurs
  • They demonstrate the process can meet defined parameters under conditions that challenge heat penetration and uniformity
  • They replace the need for biological indicators in PQ
  • They are performed without any instrumentation to simulate manual operation

Correct Answer: They demonstrate the process can meet defined parameters under conditions that challenge heat penetration and uniformity

Leave a Comment