Principles of analytical method validation MCQs With Answer

Introduction: Analytical method validation is a fundamental part of pharmaceutical analysis for B. Pharm students, ensuring methods are reliable, accurate and fit for their intended purpose. Key validation parameters include accuracy, precision, specificity, linearity, range, limit of detection (LOD), limit of quantitation (LOQ), robustness and system suitability. Understanding ICH guidelines (such as Q2(R1)), QC acceptance criteria, calibration strategies and statistical evaluation helps in developing compliant methods for drug assay, impurity testing and stability studies. Mastery of these principles ensures consistent product quality, regulatory compliance and confident interpretation of analytical data. Now let’s test your knowledge with 30 MCQs on this topic.

Q1. What is the primary purpose of analytical method validation in pharmaceuticals?

  • To check instrument ownership
  • To ensure a method is suitable, reliable and reproducible for its intended use
  • To replace method development
  • To train analysts only

Correct Answer: To ensure a method is suitable, reliable and reproducible for its intended use

Q2. Which ICH guideline is most often cited for analytical method validation?

  • ICH Q3C
  • ICH Q8
  • ICH Q2(R1)
  • ICH M7

Correct Answer: ICH Q2(R1)

Q3. Which parameter assesses the closeness of agreement between the measured value and the true value?

  • Precision
  • Accuracy
  • Robustness
  • Specificity

Correct Answer: Accuracy

Q4. Repeatability, intermediate precision and reproducibility are subcategories of which validation parameter?

  • Linearity
  • Precision
  • LOD/LOQ
  • Range

Correct Answer: Precision

Q5. Which parameter demonstrates the method’s ability to measure the analyte response proportionally across a specified concentration range?

  • Specificity
  • Linearity
  • System suitability
  • Robustness

Correct Answer: Linearity

Q6. The lowest amount of analyte that can be detected but not necessarily quantified is called:

  • Limit of Quantitation (LOQ)
  • Limit of Detection (LOD)
  • Range
  • Accuracy

Correct Answer: Limit of Detection (LOD)

Q7. Which approach is commonly used to estimate LOD and LOQ based on the calibration curve?

  • Signal-to-noise ratio only
  • Standard deviation of the response and slope (3.3σ/s and 10σ/s)
  • Visual inspection of chromatograms
  • Use of internal standard concentration

Correct Answer: Standard deviation of the response and slope (3.3σ/s and 10σ/s)

Q8. Specificity (or selectivity) in method validation refers to:

  • The method’s throughput
  • The method’s ability to measure the analyte in presence of other components like impurities, excipients and degradation products
  • The method’s cost-effectiveness
  • The reproducibility between labs

Correct Answer: The method’s ability to measure the analyte in presence of other components like impurities, excipients and degradation products

Q9. Robustness testing evaluates:

  • How the method responds to deliberate small variations in method parameters
  • Interpersonal variability only
  • The financial robustness of a lab
  • The method’s ability to detect impurities at trace levels

Correct Answer: How the method responds to deliberate small variations in method parameters

Q10. System suitability tests are performed:

  • Only after method validation is complete
  • Before sample analysis to ensure the system is working properly
  • Only for dissolution methods
  • Only for microbiological methods

Correct Answer: Before sample analysis to ensure the system is working properly

Q11. Which statistical measure is commonly used to express precision as a relative value?

  • Mean recovery
  • Standard deviation
  • Relative Standard Deviation (RSD or %RSD)
  • Correlation coefficient (r)

Correct Answer: Relative Standard Deviation (RSD or %RSD)

Q12. In accuracy studies, which experimental design is often used for recovery testing?

  • Placebo challenge
  • Standard addition at multiple concentration levels (e.g., 80%, 100%, 120%)
  • Only single concentration spike
  • System suitability injections

Correct Answer: Standard addition at multiple concentration levels (e.g., 80%, 100%, 120%)

Q13. Which of the following indicates acceptable linearity for a calibration curve in many analytical methods?

  • Correlation coefficient (r) ≤ 0.90
  • Correlation coefficient (r) ≥ 0.999 always
  • Correlation coefficient (r) typically ≥ 0.99 and residuals randomly distributed
  • RSD of standards > 10%

Correct Answer: Correlation coefficient (r) typically ≥ 0.99 and residuals randomly distributed

Q14. Which validation parameter addresses potential matrix effects in complex samples (e.g., plasma)?

  • Linearity
  • Specificity/selectivity and recovery studies
  • System suitability
  • LOD estimation

Correct Answer: Specificity/selectivity and recovery studies

Q15. What is the usual acceptance criteria for percent recovery in accuracy experiments for assay methods?

  • Recovery between 50–60%
  • Recovery typically within 98–102% for assay, depending on method and concentration
  • Recovery > 200%
  • Recovery exactly 100% only

Correct Answer: Recovery typically within 98–102% for assay, depending on method and concentration

Q16. Method transfer between laboratories requires demonstration of:

  • Only instrument brand matching
  • Equivalence of method performance (precision, accuracy, specificity) between labs
  • Different sample preparation
  • No documentation

Correct Answer: Equivalence of method performance (precision, accuracy, specificity) between labs

Q17. Which parameter defines the upper and lower concentration limits where the method is validated with acceptable accuracy and precision?

  • LOD and LOQ
  • Range
  • Specificity
  • Robustness

Correct Answer: Range

Q18. Signal-to-noise ratio method for LOQ typically uses which S/N values?

  • S/N of 1 for LOQ
  • S/N of 3 for LOQ
  • S/N of 10 for LOQ and 3 for LOD
  • S/N of 100 for LOQ

Correct Answer: S/N of 10 for LOQ and 3 for LOD

Q19. Which practice helps demonstrate specificity for chromatographic methods?

  • Running blank, placebo, standard, sample and forced degradation samples and checking peak purity
  • Only measuring a single standard
  • Using visual color comparison
  • Skipping impurity checks

Correct Answer: Running blank, placebo, standard, sample and forced degradation samples and checking peak purity

Q20. Robustness experiments commonly include deliberate changes in:

  • Ambient lighting conditions only
  • Chromatographic parameters like flow rate, pH, column temperature, mobile phase composition
  • Analyst salaries
  • Supplier invoices

Correct Answer: Chromatographic parameters like flow rate, pH, column temperature, mobile phase composition

Q21. For method precision (repeatability), which is a typical experimental requirement?

  • One replicate measurement
  • Multiple injections or analyses (e.g., six independent preparations) at a single concentration
  • Only system suitability samples
  • Only external lab data

Correct Answer: Multiple injections or analyses (e.g., six independent preparations) at a single concentration

Q22. In forced degradation studies, the main goal is to:

  • Produce market samples
  • Generate degradation products to demonstrate method specificity and stability-indicating capability
  • Reduce method sensitivity
  • Increase sample throughput

Correct Answer: Generate degradation products to demonstrate method specificity and stability-indicating capability

Q23. Which calibration approach can correct for matrix effects and recovery losses in bioanalytical methods?

  • External calibration only
  • Standard addition or use of an internal standard
  • Ignoring matrix effects
  • Using a single-point calibration without QC

Correct Answer: Standard addition or use of an internal standard

Q24. Which of the following best describes ‘system suitability’ parameters for HPLC?

  • System suitability refers to instrument warranty
  • Parameters such as theoretical plates, tailing factor, resolution and %RSD of peak areas
  • Only detector wavelength check
  • Only mobile phase pH

Correct Answer: Parameters such as theoretical plates, tailing factor, resolution and %RSD of peak areas

Q25. When validating an impurity method, what is an important acceptance criterion?

  • Impurities must not co-elute with the main peak and LOQ must be adequate to detect specification levels
  • Impurities can co-elute with the main peak
  • Only the main peak matters
  • No need to report LOQ

Correct Answer: Impurities must not co-elute with the main peak and LOQ must be adequate to detect specification levels

Q26. Which statistical test is useful to compare precision between two analysts or labs?

  • t-test for means
  • F-test for variances
  • Chi-square for proportions
  • Kruskal-Wallis for medians only

Correct Answer: F-test for variances

Q27. In validation reports, which section documents how the method was proven to be stability-indicating?

  • System suitability section only
  • Specificity/forced degradation studies section
  • Supply chain section
  • Financial summary section

Correct Answer: Specificity/forced degradation studies section

Q28. Which is a common acceptance limit for %RSD in system suitability for assay methods?

  • %RSD ≤ 20%
  • %RSD ≤ 2% for replicate injections, depending on method
  • %RSD ≥ 50%
  • No limit is needed

Correct Answer: %RSD ≤ 2% for replicate injections, depending on method

Q29. What is the difference between verification and validation in analytical methods?

  • Verification is confirming a validated method in a new lab; validation is establishing method performance originally
  • They are identical terms with no difference
  • Verification is done only for instruments
  • Validation is optional

Correct Answer: Verification is confirming a validated method in a new lab; validation is establishing method performance originally

Q30. Which documentation is essential to include at the end of a method validation study?

  • Validation protocol, raw data, calculations, statistical analysis, conclusions and acceptance criteria
  • Only a one-line summary
  • Only the final chromatogram
  • Only the analyst’s resume

Correct Answer: Validation protocol, raw data, calculations, statistical analysis, conclusions and acceptance criteria

Leave a Comment