Calibration of Gas Chromatograph (GC) is a critical skill for B. Pharm students involved in pharmaceutical analysis. This introduction covers key concepts like GC calibration, calibration curve, response factor, internal and external standards, linearity, LOD, LOQ, and system suitability. Proper calibration ensures accurate, precise, and traceable quantitative results by relating detector response to known concentrations, addressing matrix effects, and controlling instrument drift. Understanding standard preparation, verification frequency, and common pitfalls (e.g., detector saturation, carryover) is essential for reliable assay performance in pharmacopeial and regulatory contexts. Now let’s test your knowledge with 30 MCQs on this topic.
Q1. What is the primary purpose of calibrating a gas chromatograph for quantitative analysis?
- To clean the GC column before analysis
- To relate detector response to known concentrations for accurate quantitation
- To reduce carrier gas consumption
- To adjust injector temperature only
Correct Answer: To relate detector response to known concentrations for accurate quantitation
Q2. Which statement best describes an internal standard in GC calibration?
- A compound added only to the calibration standards
- A compound that is naturally present in all samples
- A compound added in a constant amount to all samples and standards to correct for variability
- A solvent used to dilute standards
Correct Answer: A compound added in a constant amount to all samples and standards to correct for variability
Q3. How is the relative response factor (RRF) defined?
- The ratio of retention times between analyte and standard
- The ratio of detector response per unit concentration of analyte to that of the standard
- The difference in peak heights of two analytes
- The ratio of solvent peak area to analyte area
Correct Answer: The ratio of detector response per unit concentration of analyte to that of the standard
Q4. On a calibration plot for GC quantitation, which axis typically represents concentration?
- Y-axis (detector response)
- X-axis (concentration)
- Z-axis (temperature)
- Legend area
Correct Answer: X-axis (concentration)
Q5. Limit of detection (LOD) in GC calibration is best described as:
- The concentration where the calibration curve becomes non-linear
- The lowest concentration that can be distinguished from background noise
- The concentration with 100% recovery
- The highest measurable concentration without dilution
Correct Answer: The lowest concentration that can be distinguished from background noise
Q6. What is the practical definition of limit of quantitation (LOQ) in GC method calibration?
- The lowest concentration that can be measured with acceptable precision and accuracy
- The concentration at which peaks overlap
- The highest calibrant used in the curve
- The detector’s maximum linear response
Correct Answer: The lowest concentration that can be measured with acceptable precision and accuracy
Q7. Which statistical parameter is commonly used to evaluate linearity of a GC calibration curve?
- Retention index
- Resolution
- Correlation coefficient (r or r²) from linear regression
- Carrier gas pressure
Correct Answer: Correlation coefficient (r or r²) from linear regression
Q8. Which of the following is NOT a typical system suitability criterion for GC quantitative analysis?
- Theoretical plates (efficiency)
- Resolution between critical peaks
- Repeatability of retention time and peak area (RSD%)
- Mass spectrometer ion source voltage
Correct Answer: Mass spectrometer ion source voltage
Q9. Which preparation approach is considered most accurate for making calibration standards?
- Volumetric dilution by eye
- Gravimetric preparation using accurate weighing and dilution
- Using pipette tips without calibration
- Preparing standards directly in the injector
Correct Answer: Gravimetric preparation using accurate weighing and dilution
Q10. When is it appropriate to use relative response factor (RRF) instead of absolute response factor?
- When no internal standard is used
- When an internal standard method is applied to correct for variability
- Only for qualitative analysis
- Only when using TCD detector
Correct Answer: When an internal standard method is applied to correct for variability
Q11. How does using a high split ratio during injection affect GC calibration?
- Increases sensitivity and lowers LOD
- Reduces the amount of analyte entering the column, which can decrease sensitivity
- Has no effect on detector response
- Makes peaks wider and increases retention times drastically
Correct Answer: Reduces the amount of analyte entering the column, which can decrease sensitivity
Q12. Detector saturation in GC calibration typically leads to what effect on the calibration curve?
- Improved linearity at high concentrations
- Deviation from linearity at high concentrations due to plateauing response
- Shorter retention times
- A change in carrier gas composition
Correct Answer: Deviation from linearity at high concentrations due to plateauing response
Q13. Why is matrix-matched calibration recommended in some pharmaceutical GC assays?
- To increase column lifetime
- To correct for matrix effects that alter detector response or extraction efficiency
- To reduce the number of calibration points
- To speed up run times
Correct Answer: To correct for matrix effects that alter detector response or extraction efficiency
Q14. Best practice for storing calibration standards is to:
- Leave them open at room temperature
- Store in amber glass vials, tightly capped, under recommended temperature conditions
- Keep them in plastic bags exposed to light
- Freeze and thaw repeatedly before use
Correct Answer: Store in amber glass vials, tightly capped, under recommended temperature conditions
Q15. When should calibration verification be performed on a GC used for routine pharmaceutical assays?
- Only once during instrument installation
- Regularly, after major maintenance, and before critical sample batches
- Never, if the instrument looks clean
- Only when changing carrier gas
Correct Answer: Regularly, after major maintenance, and before critical sample batches
Q16. Which property is most important when choosing an internal standard for GC calibration?
- It must co-elute exactly with the analyte
- It must be chemically similar in behavior but not present in the sample and must not co-elute
- It should have the highest detector response
- It must be the cheapest available compound
Correct Answer: It must be chemically similar in behavior but not present in the sample and must not co-elute
Q17. For qualitative identification using GC, calibration primarily ensures:
- That peak areas are maximal
- Retention times are comparable to reference standards and retention time locking is maintained
- That all analytes are volatile
- That detector noise is eliminated
Correct Answer: Retention times are comparable to reference standards and retention time locking is maintained
Q18. Why apply weighted regression (e.g., 1/x) to a GC calibration curve?
- To ignore low-concentration points
- To compensate for heteroscedasticity where variance increases with concentration
- To force the intercept to zero
- To reduce analysis time
Correct Answer: To compensate for heteroscedasticity where variance increases with concentration
Q19. Which measurement is generally preferred for GC quantitative analysis and why?
- Peak height, because it is less affected by baseline noise
- Peak area, because it accounts for peak shape and provides better accuracy and precision for broad or asymmetric peaks
- Retention time, because it indicates concentration
- Column temperature, because it correlates with analyte mass
Correct Answer: Peak area, because it accounts for peak shape and provides better accuracy and precision for broad or asymmetric peaks
Q20. Why are multiple calibration levels (e.g., five or more) preferred in GC calibration?
- To waste more standards
- To define the analytical range, assess linearity, and detect non-linearity across concentrations
- To increase run time unnecessarily
- To avoid using internal standards
Correct Answer: To define the analytical range, assess linearity, and detect non-linearity across concentrations
Q21. What is calibration drift in GC systems?
- A sudden, permanent change in instrument firmware
- A gradual change in instrument response or retention time over time, affecting accuracy
- The expected variation in peak area between injections
- A method to improve calibration stability
Correct Answer: A gradual change in instrument response or retention time over time, affecting accuracy
Q22. Which regulatory guideline is commonly referred to for validation of chromatographic methods including calibration?
- ICH Q2(R1)
- ISO 9001
- GMP Annex 1
- PAS 99
Correct Answer: ICH Q2(R1)
Q23. Which practice helps prevent sample carryover that could affect calibration and quantitation?
- Increasing oven temperature between runs only
- Implementing adequate injector needle wash and running blank injections
- Using the same vial for all injections
- Reducing carrier gas flow to zero
Correct Answer: Implementing adequate injector needle wash and running blank injections
Q24. What is the purpose of bracketing calibration in a long analytical run?
- To double the number of calibration standards used
- To run calibration standards before and after a batch of samples to monitor and correct for drift
- To change columns automatically
- To increase detector sensitivity mid-run
Correct Answer: To run calibration standards before and after a batch of samples to monitor and correct for drift
Q25. Which GC detector is considered most “universal” for permanent gases and inorganic components?
- Flame Ionization Detector (FID)
- Mass Spectrometer (MS)
- Thermal Conductivity Detector (TCD)
- Electron Capture Detector (ECD)
Correct Answer: Thermal Conductivity Detector (TCD)
Q26. Which formula correctly represents the Relative Response Factor (RRF) using areas and concentrations?
- RRF = (Area_std / Conc_std) / (Area_analyte / Conc_analyte)
- RRF = (Area_analyte / Conc_analyte) / (Area_std / Conc_std)
- RRF = Area_analyte × Area_std
- RRF = Conc_analyte – Conc_std
Correct Answer: RRF = (Area_analyte / Conc_analyte) / (Area_std / Conc_std)
Q27. For regulatory pharmaceutical assays, an acceptable coefficient of determination (r²) for a calibration curve is commonly:
- Less than 0.900
- Between 0.950 and 0.970
- Typically greater than 0.995, depending on method requirements
- Exactly zero
Correct Answer: Typically greater than 0.995, depending on method requirements
Q28. What is the main purpose of running a solvent blank during a calibration sequence?
- To increase sample throughput
- To check for contamination, carryover, or background peaks that could bias calibration
- To recalibrate the injection port temperature
- To flush the carrier gas line only
Correct Answer: To check for contamination, carryover, or background peaks that could bias calibration
Q29. When is the standard addition method preferred over external calibration in GC quantitation?
- When analyte concentrations are extremely high only
- When matrix effects significantly suppress or enhance analyte response and need compensation
- When no internal standard is available and matrix is simple
- When calibration curves are always linear
Correct Answer: When matrix effects significantly suppress or enhance analyte response and need compensation
Q30. How can column bleed affect calibration quality in GC analysis?
- It reduces carrier gas purity only
- It increases baseline noise and can reduce sensitivity and accuracy of low-level calibration points
- It improves peak shape for late-eluting analytes
- It shortens analysis time
Correct Answer: It increases baseline noise and can reduce sensitivity and accuracy of low-level calibration points

I am a Registered Pharmacist under the Pharmacy Act, 1948, and the founder of PharmacyFreak.com. I hold a Bachelor of Pharmacy degree from Rungta College of Pharmaceutical Science and Research. With a strong academic foundation and practical knowledge, I am committed to providing accurate, easy-to-understand content to support pharmacy students and professionals. My aim is to make complex pharmaceutical concepts accessible and useful for real-world application.
Mail- Sachin@pharmacyfreak.com

